Super User Specifications MCM 10 Conductivity ranges: 6 ranges from 2 μS/cm to 0.2 S/cm Conductivity resolution: 0.05% of dynamic range Conductivity accuracy: < 4% of the 2 μS/cm // < 2% of the 20 μS/cm range and 0.2 S/cm range // < 1% of all other ranges Acquisition time: < 500 ms/channel each frequency Frequency range (FRA): 100 Hz – 50 kHz External Dimension: 290 x 340 x 63 mm (W x D x H) Weight: 1 kg Computer connection: USB 2.0 AC input: 100-240 VAC, 50 – 60 Hz < 15 W WTSH 10 Heating/cooling type: air cooled Peltier cells stack Temperature range: -40 °C to +150 °C Temperature Accuracy: +/-0.2% (typ.) +/-0.1 °C Temperature matching: +/-0.2 °C sample to sample Number of samples:10 slots for 12 mm vials Auxiliary input: PT1000 temperature sensor External Dimension: 290 x 340 x 375 mm (W x D x H) Weight: 6 kg Computer connection: USB 2.0 AC input: 90 – 260 VAC, 50 – 60 Hz < 250 W HTCC Cell type: Parallel plates platinized platinum on glass holder Connections: 2-wire (l = 70 cm each) Cell factor: K = 1 (nominal) +/- 5% Minimal sample volume: 0.5 – 1.0 mL Temperature range: -50 °C to +180 °C Temperature ramp (max): 1 °C/min (to avoid platinum detachment)